[2026] CGH Contributions to Test Uncertainty

At AOM, we prioritize accurate metrology, and a robust uncertainty analysis is crucial for dependable measurements. For a surface figure error (SFE) test, AOM separates the contributions to the total test uncertainty into three groups: those arising from the computer-generated hologram (CGH Uncertainty), those from misalignment of the CGH to the unit under test (Alignment Uncertainty), and those from the test setup (Measurement Uncertainty). This article describes the sources of CGH Uncertainty and explains the analytical methods used to determine the corresponding contribution to the total test uncertainty. Every CGH receives a complete uncertainty profile that quantifies its single-pass RMS wavefront error uncertainty (equivalent to its contribution to the uncertainty of an SFE measurement). When the uncertainty cannot be explicitly determined, we adopt a conservative approximation. This ensures that the uncertainty reported for our CGHs is highly reliable.
Definitions
Error is defined as the deviation between measurement and true value. Of course, if the error is precisely know, it could be removed from the measurement to get the true value. In practice, the exact error is unknowable, but the uncertainty, the range within which an error value is likely to fall, can be estimated. For instance, an uncertainty of 5.0 nm RMS means that the error value is expected to fall within 5.0 nm RMS.
Wavefront error uncertainties are often computed in relation to other system errors (or uncertainties) using a sensitivity analysis. Sensitivity is a coefficient that quantifies how the test wavefront responds to specific perturbations in the system. Sensitivities for a given CGH are design-dependent and calculated using the optical model. For example, the sensitivity due to substrate thickness deviation would be reported as 0.2 nm RMS / 0.01 mm. If the thickness uncertainty is 0.005 mm, then the wavefront uncertainty due to thickness uncertainty is the product of the sensitivity and uncertainty, or 0.1 nm RMS.
For some sources of error, AOM specifies a tolerance, or an upper bound, on the uncertainty. For example, while AOM doesn’t specifically measure the substrate refractive index, it is toleranced to 0.0001 of the nominal value.
These error sources described in this article are assumed to be independent; when adding the effects of error sources reported as an RMS, the contributions should be combined as a root-sum-squared (RSS) addition. This quantity is the uncertainty stack-up.
CGH Uncertainty Contributions
Design contributions
CGH design contributions to uncertainty are determined during the design phase. They are known and quantified prior to fabrication and can be addressed in the design if they dominate the uncertainty stack-up.
- Design residual error: AOM CGHs are designed using a numerical optimization process to create a phase function matching the prescribed test surface or subsystem. This may leave a small residual wavefront error. This error is design-dependent and is given by the optical model after the optimization process. Typical values are 0.0–0.3 nm RMS.
- Encoding residual error: To prepare a CGH design for lithography, the smooth, curved features of the grating must be approximated as polygons, or “fractured”, as shown below. This approximation produces a high-spatial-frequency error in the wavefront measurement that depends on the local CGH phase slope and curvature, as well as the interferometer cutoff frequency. AOM compares the deviations of the fractured grating to the ideal grating and determines the effect on the wavefront using the average grating period and resolution on the interferometer detector. Encoding uncertainty values are typically 0.0-1.0 nm RMS. If this is a dominant form of uncertainty, customers may request finer fracturing of the design to better approximate the ideal grating. Doing so requires a larger file size and longer writing time, which increases cost.

The encoding process approximates smooth curves as discrete polygons, adding a small error to the diffracted wavefront.
- Wavelength uncertainty: Each CGH is designed for an interferometer of a specific wavelength. While the vacuum wavelength of a 632.8 nm HeNe interferometer is very stable, the wavelength in air depends on the air’s refractive index, which can vary based on factors like elevation, air pressure, humidity, and temperature. For standard measurement conditions, AOM tolerances this uncertainty to 0.002 nm and determines the corresponding wavefront error using a sensitivity from the optical model of the CGH. Typical wavefront error uncertainties are 0.0–0.5 nm RMS. When specifying a test as a customer, please note any environmental factors that may affect the interferometer wavelength. AOM can then design the CGH for the expected wavelength to reduce error in the test.
Production contributions
Production contributions consist of errors that accumulate during fabrication.
- Patterning error: To evaluate the patterning error, AOM embeds µm-scale registration marks into the CGH design, as shown below. After patterning, AOM’s lithography partners use these marks to measure the lateral errors that occurred during writing. AOM uses the RMS of the deviation magnitudes (with scale subtracted, see below) to calculate a conservative estimate of the patterning error’s contribution to wavefront uncertainty, Where λ is the test wavelength. The uncertainty values arising from patterning error are typically 0.0–3.0 nm RMS.

Note: The embedded registration marks used are on the order of the CGH grating feature size. Any effect they have on the measurement will not be resolved on standard interferometer detectors.

Left: A registration mark embedded in a CGH pattern. Right: Vector plot representing displacement errors of the registration marks.
- Pattern scale error: This contribution combines two sources of error that affect the scale of the CGH pattern. The first is the scale component of the patterning error; the patterning error contribution above captures only the noise-like component. The second is thermal expansion or contraction of the substrate, caused by any temperature difference between the pattern writing environment and the usage environment. AOM tolerances this temperature difference to 2.2 °C. Typical effects on the wavefront are 0.0–0.1 nm RMS.
Substrate contributions
The last category is contributions from the fused silica substrates. Some of the parameters are toleranced by the supplier, and some are measured after fabrication. Where applicable, the error is reported over the null pattern aperture only. Substrate size affects the magnitude of many of these contributions, with larger substrates contributing more uncertainty.
- Substrate transmitted wavefront error (TWE): The fused silica substrates used for CGHs have some as-built TWE, which carries into the wavefront of the diffracted beam. For AOM’s CGH designs, the substrate TWE in the first order is equivalent to the TWE in the zero order. AOM measures substrate TWE as part of the standard quality process before shipping; however, users can also measure it in situ. Either measurement can serve as a calibration map, and if it is subtracted from the test data, substrate TWE no longer needs to be included in the uncertainty stack-up. Calibration is recommended since uncalibrated substrate TWE is often the largest single source of error, typically 3.0–12.0 nm RMS. It is broken into “irregularity” and “power” contributions in the stack-up.
- Substrate TWE uncertainty/Substrate error measurement noise: When substrate TWE is calibrated out of the measurement, the uncertainty of the calibration map transfers to the uncertainty of the test. The primary source of uncertainty in the calibration map is measurement noise. AOM tolerances this noise to 1.3 nm RMS, repeating the measurement until the noise in the averaged result falls below this threshold. Users performing their own in situ TWE measurement should measure and tolerance the noise in the same way.
- Substrate thickness uncertainty: The thickness of the substrate is found using micrometer measurements at each of the four corners. The average of these measurements is the substrate thickness used in the CGH optical model. The variation of these measurements represents the uncertainty of the thickness. For a conservative estimate, AOM uses three times the standard deviation (3σ) for the uncertainty. Substrate thickness sensitivity is calculated from the CGH optical model. Only CGH designs using a spherical input are sensitive to substrate thickness error. Typical values of the corresponding wavefront uncertainty are <1.0 nm RMS.
- Substrate refractive index uncertainty: Error in the refractive index of the CGH substrate can cause wavefront error. AOM’s supplier tolerances the error in the refractive index to 0.0001. Only CGH designs using a spherical input are sensitive to refractive index error, and the sensitivity is derived from the optical model. Typical wavefront uncertainty values are 0.0–0.3 nm RMS.
- Substrate wedge: The CGH substrate wedge is toleranced by the supplier to one fringe (632.8 nm) per 6 in. of substrate and verified qualitatively by AOM’s lithography partners. The sensitivities of wedge in both directions (X and Y) are determined using the optical model. Typical values for the corresponding wavefront uncertainty are 0.0–0.5 nm RMS.
- Substrate bending: AOM defines substrate bending as the parallel curvature of both surfaces of the substrate (i.e., curvature that is common to both surfaces). This error is not captured in the substrate TWE map because parallel curvatures do not affect the collimation of the interferometer used to take the measurement. To calculate the error due to substrate bending, AOM first measures the SFE of the first surface of the CGH, which provides the curvature of the first surface alone. Comparing this to the power in the substrate TWE measurement gives the curvature of the second surface. The curvature common to both surfaces (i.e., the average) is the substrate bending. The corresponding wavefront error is then calculated using a sensitivity from the optical model. Typical contributions to the wavefront uncertainty from substrate bending are 0.0–3.0 nm RMS.
Reporting the Uncertainty Stack-Up
CGH documentation
For every standard CGH, the wavefront uncertainty contributions for the preceding sources of error (reported in nm RMS) can be found in the Error Analysis section of the manufacturing report. An example of the CGH Manufacture Table reporting the uncertainties is shown below. The total effect is determined by an RSS and shown as the Basic Test Uncertainty in the last line of the table. Typically, we expect a 6-inch CGH to contribute less than 5.0 nm RMS (single-pass) wavefront error to an SFE test when substrate TWE is calibrated. A 9-inch CGH typically contributes less than 10.0 nm RMS with substrate calibration.
The analysis is broken down based on whether substrate TWE is calibrated out of the measurement. For the case where substrate TWE is calibrated, the measurement uncertainty of the TWE calibration map is added to the stack-up. Otherwise, the substrate TWE is added to the stack-up. The tables quantify only the error contribution of the CGH to the test uncertainty.

Example of the CGH Manufacture Table included in the manufacturing report for each CGH.
Morpheus software
In Morpheus, the uncertainty contributions from the CGH are included on the Uncertainty page in the CGH Uncertainty section, shown in the example below. This information is contained in the Morpheus Data File (.cgh file), which is unique for each CGH and is loaded into the software before analysis. Since Morpheus automatically performs substrate TWE calibration, the Substrate Error measurement noise is included in this section, rather than the full substrate TWE.

Example of the Uncertainty page available in Morpheus.
The Morpheus Uncertainty page is designed to capture Alignment Uncertainty and Measurement Uncertainty in addition to the CGH Uncertainty to determine the complete uncertainty stack-up for the test, assuming systematic errors within the test instrument are calibrated separately. Default values for alignment contributions are prepopulated but are user-editable and can be updated based on the tolerances for a particular test configuration.
The Measurement Uncertainty section contains two sources of error. The first is the UUT (unit under test) Measurement Noise. Users should edit this field based on noise measurements or tolerances used in the test. The second is Substrate Error Registration Uncertainty, which captures the wavefront error contribution due to the uncertainty in registering the TWE calibration map to the test data using the substrate fiducials marked in each measurement. The total uncertainty stack-up for all three categories is calculated and shown at the top of the screen.
Conclusion
The errors described above are quantified for every standard AOM CGH. Other sources of CGH error in the system (e.g., etch-depth nonuniformity, duty cycle error, and refractive index nonuniformity) have been evaluated and found to be negligible. However, certain applications have tighter tolerances and require less overall test uncertainty (1–2 nm RMS); for these, AOM offers an Extreme Precision (XP) approach to CGH testing, which involves extra considerations during the design, manufacture, usage, and analysis of the test. More information about XP applications can be found in “Applications of computer-generated holograms for measuring x-ray and EUV optics” linked below.
AOM knows that precision matters, and we are continually improving our processes, measurements, and reporting to minimize error and provide you with the best possible understanding of the uncertainty of your measurement.
Further Reading
- James H. Burge, Shelby Ament, Jake Beverage, et al. “Applications of computer-generated holograms for measuring x-ray and EUV optics,” Proc. SPIE 13620, Advances in X-Ray/EUV Sources, Optics, and Components XX, 1362002 (18 Sept 2025) [link]
- Ping Zhou and James H. Burge, “Fabrication error analysis and experimental demonstration for computer-generated holograms,” Appl. Opt. 46, 657–663 (2007) [link]
- Chunyu Zhao and James H. Burge, “Optical testing with computer generated holograms: comprehensive error analysis,” Proc. SPIE 8838, Optical Manufacturing and Testing X, 88380H (7 Sept 2013) [link]
- Wenrui Cai, Ping Zhou, Chunyu Zhao, and James H. Burge, “Analysis of wavefront errors introduced by encoding computer-generated holograms,” Appl. Opt. 52, 8324–8331 (2013) [link]